VisiFerm SU RS485-ECS 225

Specification Sheet (Part/REF # 10140046-12)

Visiferm SU RS485 ECS

The VisiFerm SU RS485-ECS is designed for reliable dissolved oxygen (DO) measurement in single-use applications with the ODO Cap Sx sensor element.

Product Specifications

Sensor Family
Sensor Family
Visiferm SU
Electrical Connector
Electrical Connector
VP8
Parameter
Parameter
DO (Optical)
Sensor Output
Sensor Output
Arc: Modbus, 4-20 mA (only via Arc Wi 2G Adapter), ECS
Measuring Range
Measuring Range
4 ppb to 25 ppm (DO)
0 to 62.85 %-vol
0 to 300 %-sat
Electrolyte
Electrolyte
None
Accuracy at 25 °C
Accuracy at 25 °C
via RS485 mode
± 1.0 %-sat (at 20 %-sat), only in combination with the gamma irradiated Hamilton single-use ODO Cap and with product calibration
± 1.0 %-sat (at 100 %-sat), only in combination with the gamma irradiated Hamilton single-use ODO Cap and with product calibration
± 2.5 %-sat (at 200 %-sat), only in combination with the gamma irradiated Hamilton single-use ODO Cap and with product calibration
via ECS mode
± 2.5 %-sat (at 20 %-sat), only in combination with the gamma irradiated Hamilton single-use ODO Cap and with product calibration
± 1.0 %-sat (at 100 %-sat), only in combination with the gamma irradiated Hamilton single-use ODΟ Cap and with product calibration
± 5.0 %-sat (at 200 %-sat), only in combination with the gamma irradiated Hamilton single-use ODO Cap and with product calibration
Drift at Room Temperature
Drift at Room Temperature
< 1 % per week (of measurement value)
Oxygen Consumption
Oxygen Consumption
None
Temperature Sensor
Temperature Sensor
NTC 22 kOhm
Configurable Values
Configurable Values
Dissolved oxygen (DO): mbar, %-sat, %-vol, μg/l ppb, mg/l ppm, ppm gas; Temperature: °C, °F, K
Analog Interface
Analog Interface
ECS
Digital Interface
Digital Interface
RS485 Modbus (max. 32 addresses)
Baud Rate
Baud Rate
4800, 9600, 19200, 38400, 57600, 115200 Bd
Operating Voltage
Operating Voltage
10 to 27 VDC max. 1.5 W
Serial Number
Serial Number
Yes
Shield
Shield
No
Response Time
Response Time
t98% < 30 s at 25 °C
Certificate
Certificate
Yes, with parameter settings and materials used
ATEX Approval
ATEX Approval
No
IECEx Approval
IECEx Approval
No
Diameter
Diameter
12 mm
Autoclavable
Autoclavable
No
Steam Sterilizable
Steam Sterilizable
No
CIP
CIP
No
Gamma Irradiatable
Gamma Irradiatable
No
X-Ray irradiatable
X-Ray irradiatable
No
Operating Temperature Range
Operating Temperature Range
0 to 60 °C
Required Flow
Required Flow
None
a-length
a-length
215 mm
Spec. Version A