Hamilton's ODO Cap S3 Kit is an optical dissolved oxygen sensor cap for biopharma processes. The ODO Cap S3 Kit is designed to be mounted in single-use containers like bioreactor bags. It connects to the VisiFerm SU for reliable measurementin single-use applications.
Product Specifications
| Measurement Principle |
Measurement Principle
Fluorescence Quenching Luminophore (optical dissolved oxygen)
|
|---|---|
| Measuring Range |
Measuring Range
4 ppb to 25 ppm (DO)
0 to 62.85 %-vol 0 to 300 %-sat |
| Accuracy at 25 °C |
Accuracy at 25 °C
± 1.0 %-sat (at 20 %-sat)
± 1.0 %-sat (at 100 %-sat) ± 2.5 %-sat (at 200 %-sat) All three accuracy values are valid with calibration values provided on sensor label and after product calibration after irradiation and dry storage up to 37 months |
| Drift at 25 °C |
Drift at 25 °C
< 1 % per week (of measurement value)
|
| Wet-in Time |
Wet-in Time
< 30 min (after dry storage)
|
| Response Time at 25 °C |
Response Time at 25 °C
t98% < 30 s at 25 °C (from 0%-sat to 100%-sat)
|
| Shelf Life |
Shelf Life
24 months before irradiation
37 months after irradiation |
| Wetted Parts |
Wetted Parts
PA (Cycloaliphatic polyamide)
VMQ (Silicone elastomer) See compliance details in Material Specification document |
| Burst Pressure |
Burst Pressure
Up to 10 barg
|
| Operating Temperature Range |
Operating Temperature Range
4 to 50 °C
|
| Storage Temperature Range |
Storage Temperature Range
4 to 40 °C in non-condensing environment (<85% r. H.)
|
| Transport Temperature Range |
Transport Temperature Range
-30 to 60 °C (according to ASTM D4169-13)
|
| Sterilization |
Sterilization
Gamma irradiation compatible (up to 50 kGy)
X-Ray irradiation compatible (up to 50 kGy) Not validated for autoclavation, Sterilization In Place (SIP), nor Cleaning In Place (CIP) |
| Chemical Resistance |
Chemical Resistance
See Chemical Resistance Guide
|
Spec. Version C