The Arc Module Incyte-P in combination with the Incyte-P SU enable real-time, in-situ measurement of viable cells, making it possible to detect events and respond in real time without increased off-line sampling.
Product Specifications
| Electrical Connector |
Electrical Connector
VP8
|
|---|---|
| Parameter |
Parameter
Viable Cell Density
|
| Sensor Output |
Sensor Output
Arc: Modbus
|
| Measuring Range |
Measuring Range
0 to 700 pF/cm, equivalent to 5 x 10^5 to 8 x 10^9 cells/mL (mammalian)
|
| Temperature Sensor |
Temperature Sensor
Yes (on sensor element: Incyte-P SU)
|
| Configurable Values |
Configurable Values
VCD, Conductivity, Temperature
|
| Digital Interface |
Digital Interface
RS485 Modbus (max. 32 addresses)
|
| Baud Rate |
Baud Rate
4800, 9600, 19200, 38400, 57600, 115200 Bd
|
| Operating Voltage |
Operating Voltage
24 V (± 10 %)
|
| Serial Number |
Serial Number
Yes
|
| Certificate |
Certificate
Yes
|
| Autoclavable |
Autoclavable
No
|
| Steam Sterilizable |
Steam Sterilizable
No
|
| CIP |
CIP
No
|
| Gamma Irradiatable |
Gamma Irradiatable
No
|
| X-Ray irradiatable |
X-Ray irradiatable
No
|
| Operating Temperature Range |
Operating Temperature Range
0 to 60 °C
|
| Conductivity Range |
Conductivity Range
1 to 50 mS/cm
|
| Protection Rating |
Protection Rating
IP64 (when connected to the sensor element: Incyte-P SU)
|
Spec. Version A