VisiFerm mA 160 H3

Specification Sheet (Part/REF # 10070760-1211)

Dissolved Oxygen Sensor VisiFerm mA

Less frequent calibration and longer lifetime. 2-wire 4-20mA / HART for GMP production environments. Explosive-rated environments.

Product Specifications

Process Connection
Process Connection
PG13,5
Sensor Family
Sensor Family
VisiFerm mA
Wetted Parts
Wetted Parts
Stainless Steel 1.4435
EPDM (Ethylene propylene elastomer)
VMQ (Silicone elastomer)
See compliance details in Material Specification document
Electrical Connector
Electrical Connector
M12
Measuring Range
Measuring Range
4 ppb to 25 ppm (in liquid)
100 ppm to 600'000 ppm (in gas)
0.01 - 62.85 %-vol or 0.05 to 300 %-sat
0.1 to 625 mbar (pO2)
At 25°C and 1013 mbar total gas pressure
Electrolyte
Electrolyte
None
Accuracy at 25 °C
Accuracy at 25 °C
1 ± 0.05 %-vol; 21 ± 0.2 %-vol; 50 ± 0.5 %-vol
Drift at Room Temperature
Drift at Room Temperature
< 1 % per week
Oxygen Consumption
Oxygen Consumption
None
Temperature Sensor
Temperature Sensor
NTC 22 kOhm
Configurable Values
Configurable Values
Dissolved oxygen (DO): mbar, %-sat, %-vol, μg/l ppb, mg/l ppm, ppm gas; Temperature: °C, °F, K
Surface Quality of Steel
Surface Quality of Steel
Ra < 0.4 µm (N5)
4 to 20 mA Current Range
4 to 20 mA Current Range
3.5 to 22 mA
Analog Interface 1
Analog Interface 1
Two wire sink 4 to 20 mA for DO, programmable
4 to 20 mA Accuracy
4 to 20 mA Accuracy
< 0.3 % current value + 0.05 mA
Transmitting Power
Transmitting Power
Max. 4 dBm
Digital Interface
Digital Interface
Bluetooth 5.0
HART Protocol 7
Operating Voltage
Operating Voltage
18 to 30 VDC
Serial Number
Serial Number
Yes
Response Time
Response Time
t98% < 30 s at 25 °C
Certificate
Certificate
Yes, with measured current in air
ATEX Approval
ATEX Approval
ll 1 G Ex ia llC T3/T4/T5/T6 Ga; II 1 D Ex ia IIIC T135°C Da
IECEx Approval
IECEx Approval
ll 1 G Ex ia llC T3/T4/T5/T6 Ga; II 1 D Ex ia IIIC T135°C Da
Diameter
Diameter
12 mm
Autoclavable
Autoclavable
Yes
Steam Sterilizable
Steam Sterilizable
Yes, max. Temperature 140 °C
CIP
CIP
Yes
Operating Temperature Range
Operating Temperature Range
-20 to 140 °C; the sensor provides no DO reading above 85 °C
Pressure Range bar g
Pressure Range bar g
-1 to 12 bar
Protection Rating
Protection Rating
IP68
Required Flow
Required Flow
None
a-length
a-length
160 mm
Sensor Cap
Sensor Cap
ODO Cap H3
Spec. Version E