Conducell UPW Arc PG13,5

Conducell UPW Arc 2-pole conductivity sensors retain all the benefits of a traditional UPW sensor and is enhanced with the intelligence of Arc technology. Arc technology allows calibrations to be stored in the sensor, predictive diagnostics, automated documentation, as well as user and process assignment. Direct connection to the control system eliminates the need for a transmitter, many available communication protocols provide the flexibility for simple integration.

Found in these product categories:

Part/REF # 243579
  • (In Stock)
    (In Stock)
    Stock status is updated daily. For real-time status, please contact us.

Specifications

Sensor Family
Sensor Family
Conducell UPW
a-length
a-length
120 mm
Parameter
Parameter
Conductivity (2-pole)
Electrical Connector
Electrical Connector
VP8
Sensor Output
Sensor Output
Arc: Modbus, 4 to 20 mA
Measurement Principle
Measurement Principle
2 pole contacting
Measuring Range
Measuring Range
0.01 to 1500 µS/cm

Additional specifications

Accuracy at 25 °C
Accuracy at 25 °C
± 3 %
Cell Constant
Cell Constant
< 0.1 / cm
Temperature Sensor
Temperature Sensor
NTC 22 kOhm
Configurable Values
Configurable Values
Conductivity: µS/cm, mS/cm; Temperature: °C, °F, K
Diameter
Diameter
12 mm
Process Connection
Process Connection
PG13,5
Wetted Parts
Wetted Parts
Stainless Steel 1.4435
PEEK (Polyetheretherketone)
EPDM (Ethylene propylene elastomer)
See compliance details in Material Specification document
Surface Quality of Steel
Surface Quality of Steel
Ra < 0.4 µm (N5, electropolished)
Analog Interface 1
Analog Interface 1
4 to 20 mA for conductivity, programmable
Analog Interface 2
Analog Interface 2
4 to 20 mA for Temp., programmable
Analog Interface 1 and 2
Analog Interface 1 and 2
Galvanically not isolated; pulse width modulation 3.5 kHz
Digital Interface
Digital Interface
RS485 Modbus (max. 31 addresses)
Baud Rate
Baud Rate
4800, 9600, 19200, 38400, 57600,115200 Bd
Operating Voltage
Operating Voltage
7 to 30 VDC max. 150 mW
Serial Number
Serial Number
Yes
Certificate
Certificate
Yes, with nominal and actual cell constant
ATEX Approval
ATEX Approval
No
IECEx Approval
IECEx Approval
No
Autoclavable
Autoclavable
Yes, max. Temperature 130 °C
CIP
CIP
Yes
Steam Sterilizable
Steam Sterilizable
Yes, max. Temperature 140 °C
Operating Temperature Range
Operating Temperature Range
Analog interface: 0 to 110 °C; Digital interface: 0 to 130 °C