OneFerm pH VP 70 Pt100

The Hamilton OneFerm pH sensor is a single use glass electrode for pH measurement. It retains the high accuracy performance of a glass electrode after gamma irradiation and dry storage. It is delivered as an integrated part of a single use container to provide ease of use. The OneFerm pH sensor is only available to integrators such as bag-, bioreactor- or downstream skid manufacturers.

Found in these product categories:

Part/REF # 243216
  • (In Stock)
    (In Stock)
    Stock status is updated daily. For real-time status, please contact us.

Specifications

Sensor Family
Sensor Family
OneFerm
Parameter
Parameter
pH
a-length
a-length
70 mm
Electrical Connector
Electrical Connector
VP6
Sensor Output
Sensor Output
Traditional: mV
Measurement Principle
Measurement Principle
Combination electrode; pH potential measured against reference
Measuring Range
Measuring Range
pH 3 to 10

Additional specifications

Sensitivity
Sensitivity
Before gamma irradiation: 57 to 60 mV/pH at 25 °C (97 % -101 %); Shift after gamma irradiation: ± 1.5 mV/pH
Zero Point
Zero Point
Before gamma irradiation: ± 20 mV; Shift after gamma irradiation: ± 10 mV
Calibration Data
Calibration Data
Yes, provided on sensor label
pH Accuracy
pH Accuracy
After gamma irradiation and dry storage: ± 0.15 pH with provided calibration values. ± 0.1 pH after product calibration at the measurement temperature and within 2 pH units from the product calibration point
Drift at Room Temperature
Drift at Room Temperature
< 0.1 pH per week after wet-in time, in buffer pH 7
Membrane Shape
Membrane Shape
Spherical
Diaphragm
Diaphragm
HP-Coatramic
Temperature Sensor
Temperature Sensor
Pt100
Diameter
Diameter
12 mm
Process Connection
Process Connection
PG13,5
Wetted Parts
Wetted Parts
Glass
VMQ (Silicone elastomer)
O-ring Material
O-ring Material
FDA 21 CFR 177.2600; EG 1935/2004; USP <87>; USP <88> Class VI (121 °C)
Biocompatibility
Biocompatibility
Yes, tested according to ISO 10993-5 and USP <87>
Wet-in Time
Wet-in Time
30 min after dry storage
Shelf Life
Shelf Life
12 months prior gamma irradiation. 18 months after gamma irradiation
Autoclavable
Autoclavable
No
CIP
CIP
No
Steam Sterilizable
Steam Sterilizable
No
Gamma Irradiatable
Gamma Irradiatable
Yes, up to 45 kGy
Operating Temperature Range
Operating Temperature Range
4 to 50 °C
Sample min Conductivity
Sample min Conductivity
100 µS/cm
Pressure Range bar g
Pressure Range bar g
0 to 2 bar