Register for a Hamilton Webinar!

OxyGold B 225

OxyGold B 225 traditional dissolved oxygen sensors are well suited for applications that contain carbon dioxide or other acidic gases. The sensor offers an industry standard VP connector with a traditional nanoamp (nA) output for connection with most commercially available controllers and transmitters.

Found in these product categories:

Part/REF # 237185
  • (In Stock)
    (In Stock)
    Stock status is updated daily. For real-time status, please contact us.

Specifications

Sensor Family
Sensor Family
OxyGold B
Parameter
Parameter
DO (Electrochemical)
Sensor Output
Sensor Output
Traditional: nA
a-length
a-length
215 mm
Electrical Connector
Electrical Connector
VP6
Measurement Principle
Measurement Principle
Electrochemical reduction of oxygen
Measuring Range
Measuring Range
8 ppb to 40 ppm (DO) or 0.2 to 1000 mbar (pO₂)

Additional specifications

Drift at Room Temperature
Drift at Room Temperature
< 1 % per week
Membrane Cap
Membrane Cap
Optiflow
Electrolyte
Electrolyte
Oxylyte B
Oxygen Consumption
Oxygen Consumption
Ca. 100 ng/h in air at 25 °C
Temperature Sensor
Temperature Sensor
NTC 22 kOhm
Current in Air at 25 °C
Current in Air at 25 °C
180 to 500 nA
Diameter
Diameter
12 mm
Process Connection
Process Connection
PG13,5
Wetted Parts
Wetted Parts
Stainless Steel 1.4435
EPDM (Ethylene propylene elastomer)
VMQ (Silicone elastomer)
Silicone
See compliance details in Material Specification document
Surface Quality of Steel
Surface Quality of Steel
Ra < 0.4 µm (N5)
Electrode system
Electrode system
Silver platinum combination
Polarization Voltage and Time
Polarization Voltage and Time
0 ± 50 mV, ≥ 24 hours
Serial Number
Serial Number
Yes
Response Time t90%
Response Time t90%
< 60 s at 25 °C
Certificate
Certificate
Material and Performance Test Certificate
ATEX Approval
ATEX Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
IECEx Approval
IECEx Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
Autoclavable
Autoclavable
No
CIP
CIP
Yes
Steam Sterilizable
Steam Sterilizable
No
Operating Temperature Range
Operating Temperature Range
0 to 100 °C
Pressure Range bar g
Pressure Range bar g
0 to 12 bar
Max. CO₂ Partial Pressure
Max. CO₂ Partial Pressure
12 bar
Required Flow
Required Flow
≥ 0.1 m/s
Contact us

We're Ready to Help

Hamilton's sales, customer service, and technical support staff are available for questions and additional information about all products and services.

Contact Us