OxyGold B 120
OxyGold B 120 traditional dissolved oxygen sensors are well suited for applications that contain carbon dioxide or other acidic gases. The sensor offers an industry standard VP connector with a traditional nanoamp (nA) output for connection with most commercially available controllers and transmitters.
Found in these product categories:
Part/REF # 237180-
(In Stock)(In Stock)Stock status is updated daily. For real-time status, please contact us.
Specifications
Sensor Family |
Sensor Family
OxyGold B
|
---|---|
Parameter |
Parameter
DO (Electrochemical)
|
Sensor Output |
Sensor Output
Traditional: nA
|
a-length |
a-length
120 mm
|
Electrical Connector |
Electrical Connector
VP6
|
Measurement Principle |
Measurement Principle
Electrochemical reduction of oxygen
|
Measuring Range |
Measuring Range
8 ppb to 40 ppm (DO) or 0.2 to 1000 mbar (pO₂)
|
Additional specifications
Drift at Room Temperature |
Drift at Room Temperature
< 1 % per week
|
---|---|
Membrane Cap |
Membrane Cap
Optiflow
|
Electrolyte |
Electrolyte
Oxylyte B
|
Oxygen Consumption |
Oxygen Consumption
Ca. 100 ng/h in air at 25 °C
|
Temperature Sensor |
Temperature Sensor
NTC 22 kOhm
|
Current in Air at 25 °C |
Current in Air at 25 °C
180 to 500 nA
|
Diameter |
Diameter
12 mm
|
Process Connection |
Process Connection
PG13,5
|
Wetted Parts |
Wetted Parts
Stainless Steel 1.4435
EPDM (Ethylene propylene elastomer) Silicone - FDA 21 CFR 177.2600 VMQ (Silicone elastomer) See compliance details in Material Specification document |
Surface Quality of Steel |
Surface Quality of Steel
Ra < 0.4 µm (N5)
|
Electrode system |
Electrode system
Silver platinum combination
|
Polarization Voltage and Time |
Polarization Voltage and Time
0 ± 50 mV, ≥ 24 hours
|
Serial Number |
Serial Number
Yes
|
Response Time t90% |
Response Time t90%
< 60 s at 25 °C
|
Certificate |
Certificate
Material and Performance Test Certificate
|
ATEX Approval |
ATEX Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
|
IECEx Approval |
IECEx Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
|
Autoclavable |
Autoclavable
No
|
CIP |
CIP
Yes
|
Steam Sterilizable |
Steam Sterilizable
No
|
Operating Temperature Range |
Operating Temperature Range
0 to 100 °C
|
Pressure Range bar g |
Pressure Range bar g
0 to 12 bar
|
Max. CO₂ Partial Pressure |
Max. CO₂ Partial Pressure
12 bar
|
Required Flow |
Required Flow
≥ 0.1 m/s
|
Resources
Documents
-
Brochure
-
Catalogs
-
Certificates
- ATEX Certificate TÜV 03 ATEX 7005 X - Analog Sensors and Armatures (PDF)
- BSE / TSE / Jatropha Statement (PDF)
- Certificate - China RoHS Hazardous Substances in Parts (PDF)
- Certificates for Raw Material according the EN 10204-3.1.B
- Declaration of Quality for Calibration Solutions
- EU Declaration of Conformity - Analog Sensors and Armatures (ATEX) (PDF)
- IECEx Certificate - Analog Sensors and Housings (PDF)
- UKCA Declaration of Conformity - Analog Sensors and Armatures (UKEX) (PDF)
-
Material Specifications
-
Technical Drawings
-
Technical Manuals
Related Content
- Best Practices: Zero Point Oxygen Calibration Article
- How much torque is needed to install a sensor with PG13.5 threads? FAQ
- Where can I find documentation and certificates for my sensors? FAQ
- Documentation and Certificates for Sensors and Liquid Solutions Article
- Powered vs. non-powered cables Article
We're Ready to Help
Hamilton's sales, customer service, and technical support staff are available for questions and additional information about all products and services.
Contact Us