Oxysens 120
The Oxysens is an electrochemical dissolved oxygen sensor designed for applications in water, e.g. wastewater treatment, swimming pools or fish farms. It is easy to maintain, because the membrane and the electrolyte do not need to be replaced. The response time of the Oxysens is fast, it is almost independent to flow and insensitive to soiling.
Found in these product categories:
Part/REF # 237150-
(In Stock)(In Stock)Stock status is updated daily. For real-time status, please contact us.
Specifications
Sensor Family |
Sensor Family
Oxysens
|
---|---|
Parameter |
Parameter
DO (Electrochemical)
|
Sensor Output |
Sensor Output
Traditional: nA
|
a-length |
a-length
120 mm
|
Electrical Connector |
Electrical Connector
5 m Fix Cable / Open End
|
Measurement Principle |
Measurement Principle
Electrochemical reduction of oxygen
|
Measuring Range |
Measuring Range
40 ppb to 40 ppm (DO) or 0.8 to 1000 mbar (pO₂)
|
Additional specifications
Drift at Room Temperature |
Drift at Room Temperature
< 1 % per week
|
---|---|
Membrane Cap |
Membrane Cap
Optiflow
|
Electrolyte |
Electrolyte
Oxylyte
|
Oxygen Consumption |
Oxygen Consumption
Ca. 20 ng/h in air at 25 °C
|
Temperature Sensor |
Temperature Sensor
NTC 22 kOhm
|
Current in Air at 25 °C |
Current in Air at 25 °C
40 - 80 nA
|
Diameter |
Diameter
12 mm
|
Process Connection |
Process Connection
PG13,5
|
Wetted Parts |
Wetted Parts
Stainless Steel 1.4435
PEEK (Polyetheretherketone) EPDM (Ethylene propylene elastomer) VMQ (Silicone elastomer) Silicone |
Surface Quality of Steel |
Surface Quality of Steel
Ra < 0.8 µm (N6)
|
Electrode system |
Electrode system
Silver platinum combination
|
Polarization Voltage and Time |
Polarization Voltage and Time
-670 ± 50 mV, ≥ 2 hours
|
Serial Number |
Serial Number
Yes
|
Response Time t90% |
Response Time t90%
< 60 s at 25 °C
|
Certificate |
Certificate
No
|
ATEX Approval |
ATEX Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
|
IECEx Approval |
IECEx Approval
II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
|
Autoclavable |
Autoclavable
No
|
CIP |
CIP
No
|
Steam Sterilizable |
Steam Sterilizable
No
|
Operating Temperature Range |
Operating Temperature Range
0 to 60 °C
|
Pressure Range bar g |
Pressure Range bar g
0 to 4 bar
|
Max. CO₂ Partial Pressure |
Max. CO₂ Partial Pressure
0.01 bar
|
Required Flow |
Required Flow
≥ 0.03 m/s
|
Resources
Documents
-
Brochure
-
Catalogs
-
Certificates
- ATEX Certificate TÜV 03 ATEX 7005 X - Analog Sensors and Armatures (PDF)
- BSE / TSE / Jatropha Statement (PDF)
- Certificate - China RoHS Hazardous Substances in Parts (PDF)
- Certificates for Raw Material according the EN 10204-3.1.B
- Declaration of Quality for Calibration Solutions
- EU Declaration of Conformity - Analog Sensors and Armatures (ATEX) (PDF)
- IECEx Certificate - Analog Sensors and Housings (PDF)
- UKCA Declaration of Conformity - Analog Sensors and Armatures (UKEX) (PDF)
-
Material Specifications
-
Technical Drawings
-
Technical Manuals
Related Content
- Understanding Surface Finish Article
- Best Practices: Zero Point Oxygen Calibration Article
- How much torque is needed to install a sensor with PG13.5 threads? FAQ
- Where can I find documentation and certificates for my sensors? FAQ
- Documentation and Certificates for Sensors and Liquid Solutions Article
We're Ready to Help
Hamilton's sales, customer service, and technical support staff are available for questions and additional information about all products and services.