Register for a Hamilton Webinar!

VisiFerm DO SU ECS 225

The VisiFerm DO SU ECS 225 for single-use applications works with Hamilton single-use ODO Cap . The VisiFerm DO SU ECS 225 is reusable and can be mounted directly into the ODO Cap within the bag or SU bioreactor.

Found in these product categories:

Part/REF # 10116428
  • (In Stock)
    (In Stock)
    Stock status is updated daily. For real-time status, please contact us.

Specifications

Sensor Family
Sensor Family
Visiferm SU
a-length
a-length
215 mm
Electrical Connector
Electrical Connector
VP8
Sensor Output
Sensor Output
Traditional: nA
Measurement Principle
Measurement Principle
Oxygen dependent luminescence quenching, temperature compensation
Measuring Range
Measuring Range
4 ppb to 25 ppm (DO)
0 to 62.85 %-vol
0 to 300 %-sat
Accuracy at 25 °C
Accuracy at 25 °C
± 1.0 %-sat (at 100 %-sat), only in combination with the gamma irradiated Hamilton single-use ODO Cap and with product calibration
± 2.5 %-sat (at 20 %-sat), only in combination with the gamma irradiated Hamilton single-use ODO Cap and with product calibration
± 5.0 %-sat (at 200 %-sat), only in combination with the gamma irradiated Hamilton single-use ODO Cap and with product calibration

Additional specifications

Drift at Room Temperature
Drift at Room Temperature
< 1 % per week (of measurement value)
Electrolyte
Electrolyte
None
Oxygen Consumption
Oxygen Consumption
None
Temperature Sensor
Temperature Sensor
NTC 22 kOhm
Configurable Values
Configurable Values
DO: mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm (gas); ppb/ppm (dissolved oxygen); Temperature: °C
Diameter
Diameter
12 mm
Analog Interface 1
Analog Interface 1
ECS / 4 to 20 mA
Digital Interface
Digital Interface
RS485 Modbus (max. 31 addresses)
Baud Rate
Baud Rate
4800, 9600, 19200, 38400, 57600,115200 Bd
Operating Voltage
Operating Voltage
7 to 30 VDC max. 1 W
Serial Number
Serial Number
Yes
Response Time
Response Time
t98% < 30 s at 25 °C
Certificate
Certificate
Yes, with parameter settings and materials used
ATEX Approval
ATEX Approval
No
Autoclavable
Autoclavable
No
CIP
CIP
No
Steam Sterilizable
Steam Sterilizable
No
Gamma Irradiatable
Gamma Irradiatable
No
Operating Temperature Range
Operating Temperature Range
0 to 60 °C
Required Flow
Required Flow
None