Dencytee RS485 325
Found in these product categories:
Part/REF # 10064919-13
-
(In Stock)(In Stock)Stock status is updated daily. For real-time status, please contact us.
Specifications
Sensor Family |
Sensor Family
Dencytee Arc
|
---|---|
a-length |
a-length
325 mm
|
Parameter |
Parameter
Total Cell Density
|
Electrical Connector |
Electrical Connector
VP8
|
Sensor Output |
Sensor Output
Arc: Modbus
|
Measurement Principle |
Measurement Principle
Transmission and Reflection (including temperature compensation, daylight filter and subtraction)
|
Wavelength |
Wavelength
860 nm
|
Additional specifications
Measuring Range |
Measuring Range
e.g. 0-200g/l cell dry weight yeast
0-4 AU 0-30'000 NTU |
---|---|
Accuracy at 25 °C |
Accuracy at 25 °C
Resolution: 0.05 g/l (0-10 g/l); 0.5 % (> 10 g/l) cell dry weight yeast
Reproducibility: ± 1 % Linearity: < ± 1% of respective measuring range (specific to application with off-line correlation) |
Temperature Sensor |
Temperature Sensor
NTC 22 kOhm
|
Configurable Values |
Configurable Values
TCD: PCV, AU, arb. Unit, NTU, g/l, CFU, e6 cells/ml, OD
Temperature: K, °C, °F |
Diameter |
Diameter
12 mm
|
Process Connection |
Process Connection
PG13,5
|
Wetted Parts |
Wetted Parts
Stainless Steel 1.4435
Sapphire glass EPDM (Ethylene propylene elastomer) See compliance details in Material Specification document |
Surface Quality of Steel |
Surface Quality of Steel
Ra < 0.4 µm (N5)
|
Digital Interface |
Digital Interface
RS485 Modbus (max. 31 addresses)
|
Baud Rate |
Baud Rate
4800, 9600, 19200, 38400, 57600,115200 Bd
|
Operating Voltage |
Operating Voltage
+10 to +27 VDC
|
Optical Path Length |
Optical Path Length
5 mm
|
Serial Number |
Serial Number
Yes
|
Certificate |
Certificate
Yes, in water and at 10'000 NTU Standard
|
ATEX Approval |
ATEX Approval
No
|
IECEx Approval |
IECEx Approval
No
|
Autoclavable |
Autoclavable
Yes
|
CIP |
CIP
Yes
|
Steam Sterilizable |
Steam Sterilizable
Yes, max. Temperature 140 °C
|
Storage Conditions |
Storage Conditions
-20 °C to 45 °C
|
Operating Temperature Range |
Operating Temperature Range
0 to 140 °C; the sensor provides no TCD reading above 80 °C.
|
Pressure Range bar g |
Pressure Range bar g
0 to 12 bar
|
Protection Rating |
Protection Rating
IP 68
|
Resources
Documents
-
Brochure
-
Catalogs
-
Certificates
-
Flyers
-
Material Specifications
-
Technical Drawings
-
Technical Manuals
Related Content
- Which wavelength is used for TCD measurement? FAQ
- What is special about the Dencytee Arc? FAQ
- When is a replacement of the Dencytee Arc Sensor recommended? FAQ
- At what value of the quality indicator does the Dencytee Arc need to be replaced? FAQ
- Is the measurement from Dencytee Arc less reliable if the quality index decreases? FAQ

We're Ready to Help
Hamilton's sales, customer service, and technical support staff are available for questions and additional information about all products and services.
Contact Us