Register for a Hamilton Webinar!

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 9.21 DuraCal liquid buffer solutions have an accuracy of +/- 0.02 pH over 60 months. Like all aqueous solutions, the pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.

pH vs. Temperature: Reference table for 9.21 DuraCal pH buffer

(Refs 238319, 238219, 238919, 238216, 238897)

Temp [°C]pH ValueTemp [°C]pH Value
5 9.45 30 9.16
6 9.43 31 9.15
7 9.42 32 9.14
8 9.41 33 9.13
9 9.39 34 9.12
10 9.38 35 9.11
11 9.37 36 9.10
12 9.36 37 9.09
13 9.34 38 9.08
14 9.33 39 9.07
15 9.32 40 9.06
16 9.31 41 9.06
17 9.30 42 9.05
18 9.28 43 9.04
19 9.27 44 9.03
20 9.26 45 9.03
21 9.25 46 9.02
22 9.24 47 9.01
23 9.23 48 9.01
24 9.22 49 9.00
25 9.21 50 8.99
26 9.20
27 9.19
28 9.18
29 9.17