Register for a Hamilton Webinar!

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 8.00 DuraCal liquid buffer solution (Ref 238277) has an accuracy of +/- 0.02 pH units and a stability of 60 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.

pH vs. Temperature: Reference table for 8.00 DuraCal pH buffer

Temp [°C]pH ValueTemp [°C]pH Value
5 8.09 30 7.98
6 8.09 31 7.98
7 8.08 32 7.98
8 8.08 33 7.97
9 8.07 34 7.97
10 8.07 35 7.97
11 8.06 36 7.96
12 8.06 37 7.96
13 8.05 38 7.96
14 8.05 39 7.95
15 8.05 40 7.95
16 8.04 41 7.95
17 8.04 42 7.95
18 8.03 43 7.94
19 8.03 44 7.94
20 8.02 45 7.94
21 8.02 46 7.94
22 8.01 47 7.94
23 8.01 48 7.93
24 8.01 49 7.93
25 8.00 50 7.93
26 8.00
27 7.99
28 7.99
29 7.99