Temperature Influence on DuraCal pH Buffer 8.00 (Ref 238277)
A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 8.00 DuraCal liquid buffer solution (Ref 238277) has an accuracy of +/- 0.02 pH units and a stability of 60 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.
pH vs. Temperature: Reference table for 8.00 DuraCal pH buffer
Temp [°C] | pH Value | Temp [°C] | pH Value |
---|---|---|---|
5 | 8.09 | 30 | 7.98 |
6 | 8.09 | 31 | 7.98 |
7 | 8.08 | 32 | 7.98 |
8 | 8.08 | 33 | 7.97 |
9 | 8.07 | 34 | 7.97 |
10 | 8.07 | 35 | 7.97 |
11 | 8.06 | 36 | 7.96 |
12 | 8.06 | 37 | 7.96 |
13 | 8.05 | 38 | 7.96 |
14 | 8.05 | 39 | 7.95 |
15 | 8.05 | 40 | 7.95 |
16 | 8.04 | 41 | 7.95 |
17 | 8.04 | 42 | 7.95 |
18 | 8.03 | 43 | 7.94 |
19 | 8.03 | 44 | 7.94 |
20 | 8.02 | 45 | 7.94 |
21 | 8.02 | 46 | 7.94 |
22 | 8.01 | 47 | 7.94 |
23 | 8.01 | 48 | 7.93 |
24 | 8.01 | 49 | 7.93 |
25 | 8.00 | 50 | 7.93 |