Register for a Hamilton Webinar!

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 7.00 DuraCal liquid buffer solutions are considered "Certified Reference Material" and are tested by a 3rd party lab - DAkkS - against primary reference solutions from PTB and NIST. 7.00 liquid buffers have an accuracy of +/- 0.01 pH over 24 months and +/- 0.02 pH over 60 months. Like all aqueous solutions, the pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.

pH vs. Temperature: Reference table for 7.00 Duracal pH buffers

(Refs 238318, 238218, 238918, 238333, 238188, 238896)

Temp [°C]pH ValueTemp [°C]pH Value
5 7.09 30 6.99
6 7.08 31 6.99
7 7.08 32 6.98
8 7.07 33 6.98
9 7.07 34 6.98
10 7.06 35 6.98
11 7.06 36 6.98
12 7.05 37 6.98
13 7.05 38 6.97
14 7.04 39 6.97
15 7.04 40 6.97
16 7.04 41 6.97
17 7.03 42 6.97
18 7.03 43 6.97
19 7.02 44 6.97
20 7.02 45 6.97
21 7.02 46 6.97
22 7.01 47 6.97
23 7.01 48 6.97
24 7.01 49 6.97
25 7.00 50 6.97
26 7.00
27 7.00
28 6.99
29 6.99