Temperature Influence on DuraCal pH Buffer 5.00 (Ref 238275)
A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 5.00 DuraCal liquid buffer solution (Ref 238275) has an accuracy of +/- 0.02 pH units and a stability of 60 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.
pH vs. Temperature: Reference table for 5.00 DuraCal pH buffer
Temp [°C] | pH Value | Temp [°C] | pH Value |
---|---|---|---|
5 | 5.04 | 30 | 5.00 |
6 | 5.03 | 31 | 5.00 |
7 | 5.03 | 32 | 5.00 |
8 | 5.03 | 33 | 5.00 |
9 | 5.03 | 34 | 5.00 |
10 | 5.02 | 35 | 5.00 |
11 | 5.02 | 36 | 5.00 |
12 | 5.02 | 37 | 5.00 |
13 | 5.02 | 38 | 5.00 |
14 | 5.01 | 39 | 5.00 |
15 | 5.01 | 40 | 5.00 |
16 | 5.01 | 41 | 5.00 |
17 | 5.01 | 42 | 5.01 |
18 | 5.01 | 43 | 5.01 |
19 | 5.01 | 44 | 5.01 |
20 | 5.01 | 45 | 5.01 |
21 | 5.00 | 46 | 5.01 |
22 | 5.00 | 47 | 5.01 |
23 | 5.00 | 48 | 5.02 |
24 | 5.00 | 49 | 5.02 |
25 | 5.00 | 50 | 5.02 |
26 | 5.00 | ||
27 | 5.00 | ||
28 | 5.00 | ||
29 | 5.00 |