Temperature Influence on DuraCal pH Buffer 3.06 (Ref 238274)
A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 3.06 DuraCal liquid buffer solution (Ref 238274) has an accuracy of +/- 0.02 pH units and a stability of 60 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.
pH vs. Temperature: Reference table for 3.06 DuraCal pH buffer
Temp [°C] | pH Value | Temp [°C] | pH Value |
---|---|---|---|
5 | 3.12 | 30 | 3.05 |
6 | 3.12 | 31 | 3.05 |
7 | 3.11 | 32 | 3.05 |
8 | 3.11 | 33 | 3.05 |
9 | 3.11 | 34 | 3.04 |
10 | 3.10 | 35 | 3.04 |
11 | 3.10 | 36 | 3.04 |
12 | 3.10 | 37 | 3.04 |
13 | 3.09 | 38 | 3.04 |
14 | 3.09 | 39 | 3.04 |
15 | 3.09 | 40 | 3.04 |
16 | 3.08 | 41 | 3.04 |
17 | 3.08 | 42 | 3.04 |
18 | 3.08 | 43 | 3.04 |
19 | 3.08 | 44 | 3.04 |
20 | 3.07 | 45 | 3.04 |
21 | 3.07 | 46 | 3.04 |
22 | 3.07 | 47 | 3.04 |
23 | 3.07 | 48 | 3.04 |
24 | 3.06 | 49 | 3.04 |
25 | 3.06 | 50 | 3.04 |
26 | 3.06 | ||
27 | 3.06 | ||
28 | 3.05 | ||
29 | 3.05 |