Temperature Influence on DuraCal pH Buffer 2.00 (Ref 238273)
A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 2.00 DuraCal liquid buffer solution (Ref 238273) has an accuracy of +/- 0.02 pH units and a stability of 60 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.
pH vs. Temperature: Reference table for 2.00 DuraCal pH buffer
Temp [°C] | pH Value | Temp [°C] | pH Value |
---|---|---|---|
5 | 1.99 | 30 | 1.99 |
6 | 1.99 | 31 | 1.99 |
7 | 1.99 | 32 | 1.99 |
8 | 2.00 | 33 | 1.99 |
9 | 2.00 | 34 | 1.98 |
10 | 2.00 | 35 | 1.98 |
11 | 2.00 | 36 | 1.98 |
12 | 2.00 | 37 | 1.98 |
13 | 2.00 | 38 | 1.98 |
14 | 2.00 | 39 | 1.98 |
15 | 2.00 | 40 | 1.98 |
16 | 2.00 | 41 | 1.98 |
17 | 2.00 | 42 | 1.97 |
18 | 2.00 | 43 | 1.97 |
19 | 2.00 | 44 | 1.97 |
20 | 2.00 | 45 | 1.97 |
21 | 2.00 | 46 | 1.97 |
22 | 2.00 | 47 | 1.97 |
23 | 2.00 | 48 | 1.97 |
24 | 2.00 | 49 | 1.97 |
25 | 2.00 | 50 | 1.97 |
26 | 2.00 | ||
27 | 1.99 | ||
28 | 1.99 | ||
29 | 1.99 |