Register for a Hamilton Webinar!

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 2.00 DuraCal liquid buffer solution (Ref 238273) has an accuracy of +/- 0.02 pH units and a stability of 60 months. The pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.

pH vs. Temperature: Reference table for 2.00 DuraCal pH buffer

Temp [°C]pH ValueTemp [°C]pH Value
5 1.99 30 1.99
6 1.99 31 1.99
7 1.99 32 1.99
8 2.00 33 1.99
9 2.00 34 1.98
10 2.00 35 1.98
11 2.00 36 1.98
12 2.00 37 1.98
13 2.00 38 1.98
14 2.00 39 1.98
15 2.00 40 1.98
16 2.00 41 1.98
17 2.00 42 1.97
18 2.00 43 1.97
19 2.00 44 1.97
20 2.00 45 1.97
21 2.00 46 1.97
22 2.00 47 1.97
23 2.00 48 1.97
24 2.00 49 1.97
25 2.00 50 1.97
26 2.00
27 1.99
28 1.99
29 1.99