Temperature Influence on DuraCal pH Buffer 10.01
A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 10.01 DuraCal liquid buffer solutions are considered "Certified Reference Material" and are tested by a 3rd party lab - DAkkS - against primary reference solutions from PTB and NIST. 10.01 liquid buffers have an accuracy of +/- 0.02 pH over 60 months. Like all aqueous solutions, the pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.
pH vs. Temperature: Reference table for 10.01 DuraCal buffer
(Refs 238321, 238223, 238923, 238187, 238898)
Temp [°C] | pH Value | Temp [°C] | pH Value |
---|---|---|---|
5 | 10.19 | 30 | 9.97 |
6 | 10.18 | 31 | 9.95 |
7 | 10.17 | 32 | 9.94 |
8 | 10.17 | 33 | 9.93 |
9 | 10.16 | 34 | 9.92 |
10 | 10.15 | 35 | 9.92 |
11 | 10.14 | 36 | 9.91 |
12 | 10.13 | 37 | 9.90 |
13 | 10.13 | 38 | 9.89 |
14 | 10.12 | 39 | 9.88 |
15 | 10.11 | 40 | 9.86 |
16 | 10.10 | 41 | 9.86 |
17 | 10.09 | 42 | 9.85 |
18 | 10.08 | 43 | 9.84 |
19 | 10.07 | 44 | 9.83 |
20 | 10.06 | 45 | 9.83 |
21 | 10.05 | 46 | 9.82 |
22 | 10.04 | 47 | 9.81 |
23 | 10.03 | 48 | 9.80 |
24 | 10.02 | 49 | 9.80 |
25 | 10.01 | 50 | 9.79 |
26 | 10.00 | ||
27 | 9.99 | ||
28 | 9.98 | ||
29 | 9.97 |