Register for a Hamilton Webinar!

A prerequisite for an accurate pH calibration is the availability of suitable buffer solutions. Hamilton 10.01 DuraCal liquid buffer solutions are considered "Certified Reference Material" and are tested by a 3rd party lab - DAkkS - against primary reference solutions from PTB and NIST. 10.01 liquid buffers have an accuracy of +/- 0.02 pH over 60 months. Like all aqueous solutions, the pH value of the buffer is affected by changing temperature. The effects of temperature from 5 to 50°C is shown below in 1 degree increments.

pH vs. Temperature: Reference table for 10.01 DuraCal buffer

(Refs 238321, 238223, 238923, 238187, 238898)

Temp [°C]pH ValueTemp [°C]pH Value
5 10.19 30 9.97
6 10.18 31 9.95
7 10.17 32 9.94
8 10.17 33 9.93
9 10.16 34 9.92
10 10.15 35 9.92
11 10.14 36 9.91
12 10.13 37 9.90
13 10.13 38 9.89
14 10.12 39 9.88
15 10.11 40 9.86
16 10.10 41 9.86
17 10.09 42 9.85
18 10.08 43 9.84
19 10.07 44 9.83
20 10.06 45 9.83
21 10.05 46 9.82
22 10.04 47 9.81
23 10.03 48 9.80
24 10.02 49 9.80
25 10.01 50 9.79
26 10.00
27 9.99
28 9.98
29 9.97