Temperature Influence of Conductivity Standard 147 µS/cm (Ref 238985)
Hamilton offers the 147 µS/cm liquid conductivity standard (Ref 238985), whose stability of ±1% is guaranteed over a lifecycle of up to 18 months. These standards can be used several times, on condition that the bottle is not left open without its lid for more than one hour total. The composition of these standards is patented, and the procedure for determining conductivity has been developed in collaboration with the Danish Institute of Fundamental Metrology (FDM) and the National Institute of Standards and Technology (NIST). Please refer to the calibration certificate packaged with every HAMILTON Conductivity Standard for further details.
Conductivity vs. Temperature: Reference table
Temperature °C | Conductivity µS/cm | Temperature °C | Conductivity µS/cm | Temperature °C | Conductivity µS/cm |
---|---|---|---|---|---|
5 | 92 | 21 | 136 | 37 | 183 |
6 | 95 | 22 | 138 | 38 | 186 |
7 | 97 | 23 | 141 | 39 | 189 |
8 | 100 | 24 | 144 | 40 | 192 |
9 | 103 | 25 | 147 | 41 | 195 |
10 | 105 | 26 | 150 | 42 | 199 |
11 | 108 | 27 | 153 | 43 | 202 |
12 | 111 | 28 | 156 | 44 | 205 |
13 | 113 | 29 | 159 | 45 | 209 |
14 | 116 | 30 | 162 | 46 | 212 |
15 | 119 | 31 | 165 | 47 | 215 |
16 | 122 | 32 | 167 | 48 | 218 |
17 | 124 | 33 | 171 | 49 | 222 |
18 | 127 | 34 | 174 | 50 | 225 |
19 | 130 | 35 | 176 | ||
20 | 133 | 36 | 180 |