Hamilton offers the 1413 µS/cm liquid conductivity standard in a convenient Calpack plastic bottle with stability of ±1% guaranteed over a lifecycle of up to 18 months. These standards can be used several times, on condition that the bottle is not left open without its lid for more than one hour total. The composition of these standards is patented, and the procedure for determining conductivity has been developed in collaboration with the Danish Institute of Fundamental Metrology (FDM) and the National Institute of Standards and Technology (NIST). Please refer to the calibration certificate packaged with every HAMILTON Conductivity Standard for further details.

Conductivity vs. Temperature: Reference table for 1413 μS/cm (Plastic Bottle)

Temperature °C
Conductivity µS/cm
Temperature °C
Conductivity µS/cm
Temperature °C
Conductivity µS/cm
5 893 21 1305 37 1751
6 918 22 1329 38 1780
7 943 23 1358 39 1810
8 968 24 1385 40 1843
9 993 25 1413 41 1869
10 1018 26 1440 42 1899
11 1044 27 1468 43 1929
12 1069 28 1495 44 1959
13 1095 29 1523 45 1988
14 1121 30 1552 46 2020
15 1149 31 1579 47 2051
16 1173 32 1607 48 2082
17 1199 33 1636 49 2113
18 1225 34 1665 50 2145
19 1251 35 1691
20 1278 36 1722