Temperature Influence of Conductivity Standard 1413 µS/cm (Ref 238928 Glass Bottle)
Hamilton offers the 1413 µS/cm liquid conductivity standard in a glass bottle, whose stability of ±1% is guaranteed over a lifecycle of up to 3 years. These standards can be used several times, on condition that the glass bottle is not left open without its lid for more than one hour total. The composition of these standards is patented, and the procedure for determining conductivity has been developed in collaboration with the Danish Institute of Fundamental Metrology (FDM) and the National Institute of Standards and Technology (NIST). Please refer to the calibration certificate packaged with every Hamilton Conductivity Standard for further details.
Conductivity vs. Temperature: Reference table for 1413 μS/cm (Glass Bottle)
Temperature °C | Conductivity µS/cm | Temperature °C | Conductivity µS/cm | Temperature °C | Conductivity µS/cm |
---|---|---|---|---|---|
5 | 717 | 21 | 1255 | 37 | 1944 |
6 | 746 | 22 | 1293 | 38 | 1992 |
7 | 775 | 23 | 1333 | 39 | 2039 |
8 | 805 | 24 | 1373 | 40 | 2088 |
9 | 836 | 25 | 1413 | 41 | 2137 |
10 | 868 | 26 | 1455 | 42 | 2186 |
11 | 899 | 27 | 1497 | 43 | 2235 |
12 | 932 | 28 | 1539 | 44 | 2285 |
13 | 965 | 29 | 1582 | 45 | 2336 |
14 | 999 | 30 | 1626 | 46 | 2387 |
15 | 1033 | 31 | 1670 | 47 | 2438 |
16 | 1069 | 32 | 1714 | 48 | 2490 |
17 | 1105 | 33 | 1759 | 49 | 2542 |
18 | 1142 | 34 | 1805 | 50 | 2594 |
19 | 1179 | 35 | 1851 | ||
20 | 1217 | 36 | 1897 |