Register for a Hamilton Webinar!

ProductEthylene OxideAutoclaveChemical1
Cemented Needle (N)
Special Cemented Needle (SN)
Yes No Yes, no bleach
Luer Tip Cemented Needle (LTN)
Luer Tip Special Cemented Needle (LTSN)
Yes No Yes, no bleach
Luer Tip (LT) Yes Yes Yes
Removable Needle, (RN) Yes Yes Yes
Knurled Hub (KH) Yes Yes Yes
Fixed Needle (FN) Yes No Yes, no bleach
PTFE Luer Lock (TLL) Yes Yes, except 25 mL and larger Yes
Bubble Free Prime (BFP) Yes No Yes
SampleLock (SL) Yes No Yes, no bleach
ChemSeal (C)
ChemSeal with Stop (CX)
Yes Yes Yes
Carbon Analyzer (CA) Yes No Yes, no bleach
Diluter (D)
Diluter with Stop (DX)
Yes Yes, except 25 mL and larger Yes
AccuDil (AD)
Diluter AccuDil (DAD)
Yes No Yes
Constant Rate (CR) Yes No Yes, no bleach
Removable Needles (RN) (18-26s gauge) Yes Yes Yes
Removable Needles (RN) (27-34 gauge) Yes Yes Yes
Metal Hub Needles (N) Yes Yes Yes
Kel-F Hub Needles (KF) Yes Yes Yes
PTFE Tubing Yes Yes Yes
Kel-F Fittings Yes Yes Yes

Notes:

  • For syringes that can be autoclaved, they need to be disassembled prior to autoclaving.
  • 1Recommended disinfecting chemicals are Microcide SQ ® (p/n 3995-01), 10% bleach, acetone or ethanol.