Register for a Hamilton Webinar!

The below table shows all the recommended maximum dilutions for every possible volume pairing of Microlab 600 syringes.

Right-Syringe Volume in Top Row; Left-Syringe Volume in Left Column

10 µL25 µL50 µL100 µL250 µL500 µL1000 µL2500 µL5000 µL10000 µL25000 µL50000 µL
10 µL 1:10
25 µL 1:25 1:10
50 µL 1:50 1:20 1:10
100 µL 1:100 1:40 1:20 1:10
250 µL 1:250 1:100 1:50 1:25 1:10
500 µL 1:500 1:200 1:100 1:50 1:20 1:10
1000 µL 1:1000 1:400 1:200 1:100 1:40 1:20 1:10
2500 µL 1:2500 1:1000 1:500 1:250 1:100 1:50 1:25 1:10
5000 µL 1:5000 1:2000 1:1000 1:500 1:200 1:100 1:50 1:20 1:10
10000 µL 1:10000 1:4000 1:2000 1:1000 1:400 1:200 1:100 1:40 1:20 1:10
25000 µL 1:25000 1:10000 1:5000 1:2500 1:1000 1:500 1:250 1:100 1:50 1:25 1:10
50000 µL 1:50000 1:20000 1:10000 1:5000 1:2000 1:1000 1:500 1:200 1:100 1:50 1:20 1:10