Register for a Hamilton Webinar!

Reversed-Phase Columns

Packing NameMaterialExchange CapacityPore SizeSurface Area (m2/cm3)Pore Volume (cm3/g)Temp LimitsMobile Phase LimitsBuffer StrengthMaximum Pressure

PRP-1

PS-DVB*N/A100 Å4000.7725-85 °CpH 1-13, 0-100% aqueous, organic modifier0.0-0.5 N5,000 psi
PRP-C18C18 bonded to PS-DVB*N/A100 Å3680.6245-85 °CpH 1-13, 0-100% aqueous, organic modifier0.0-0.5 N5,000 psi
PRP-3PS-DVB*N/A
300 Å
480.205-85 °C
pH 1-13, 0-100% aqueous, organic modifier
0.0-0.5 N
5,000 psi
PRP-h5
Fluorinated PS-DVB*
N/A
300 Å
2770.5265-85 °C
pH 1-13, 0-100% aqueous, organic modifier
0.0-0.5 N
5,000 psi
HxSil C8SilicaN/A110 Å320 (m2/g)0.95-60 °CpH 1.5-10, 0-100% aqueous, organic modifier
N/A5,000 psi
HxSil C18
Silica
N/A
110 Å
320 (m2/g)0.95-60 °C
pH 1.5-10, 0-100% aqueous, organic modifier
N/A
5,000 psi

Anion Exchange Columns

Packing NameSupport MaterialExchange CapacityPore SizeSurface Area (m2/cm3)Pore Volume (cm3/g)Temp. LimitsMobile Phase LimitsBuffer StrengthMaximum Pressure
PRP-X100PS-DVB* with Trimethyl ammonium Exchanger0.19 meq/gm100 Å4050.64pH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-0.5 N5,000 psi
PRP-X110PS-DVB* with Trimethyl ammonium Exchanger0.11 meq/gm100 Å4760.728pH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-0.5 N5,000 psi
PRP-X500Poly(meth-acryl amido-propyl Trimethyl-ammonium chloride)1.6 meq/gmSuperficially porous

1.4

0.0063pH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-2.0 N5,000 psi
PRP-X600(di-methyl amido-propyl meth-acrylamide)1.6 meq/gmSuperficially porous1.40.0046pH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-1.0 N5,000 psi
RCX-10PS-DVB* with Trimethyl ammonium Exchanger0.35 meq/gm100 Å3580.624pH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-1.0 N5,000 psi
RCX-30PS-DVB* with Trimethyl ammonium Exchanger1.0 meq/gm100 Å1050.28pH Dependent**pH 1-13. 0-100% aqueous, organic modifier0.0-1.0 N5,000 psi

Cation Exchange Columns

Packing NameSupport MaterialExchange CapacityPore SizeSurface Area (m2/cm3)Pore Volume (cm3/g)Temp. LimitsMobile Phase LimitsBuffer StrengthMaximum Pressure
PRP-X200PS-DVB* Sulfonate Exchanger35 ueq/gm100 Å4500.7525-60 °CpH 1-13. 0-100% aqueous, organic modifier0.0-0.5 N5,000 psi
PRP-X400PS-DVB* Sulfonate Exchanger2.5 meq/gmN/A8.250.05765-60 °CpH 1-13. 0-100% aqueous, organic modifier0.0-5.0 N5,000 psi
PRP-X800PS-DVB* Itaconate Exchanger1.6 meq/gm100 Å500.1665-60 °C0-100% aqueous, organic modifier0.0-0.5N5,000 psi
HC-40PS-DVB* Sulfonate Exchanger5 meq/gmGel-typeNANA25-90 °C100% WaterWater400 psi
HC-75 H+PS-DVB* Sulfonate Exchanger5 meq/gmGel-typeNANA25-90 °C100% Water 0-40% Acetonitrile WaterWater400 psi
HC-75 Ca+2PS-DVB* Sulfonate Exchanger5 meq/gmGel-typeNANA25-90 °C100% Water 0-40% Acetonitrile WaterWater400 psi
HC-75 Pb+2PS-DVB* Sulfonate Exchanger5 meq/gmGel-typeNANA25-90 °C100% Water 0-40% Acetonitrile WaterWater400 psi

Ion Exclusion

Packing NameSupport MaterialExchange CapacityPore SizeSurface Area (m2/cm3)Pore Volume (cm3/g)Temp. LimitsMobile Phase LimitsBuffer StrengthMaximum Pressure
PRP-X300PS-DVB* Sulfonate Exchanger0.17 meq/gm100 Å3600.7485-60 °CpH 1-13. 0-100% aqueous, organic modifier0.0-0.5 N5,000 psi

*PS-DVB is Poly(styrene-divinylbenzene)

**(5-60 °C, pH 1-7.9) / (5-30 °C, pH 8-13)