Stop Flow Rate (Microlab STAR Liquid Editor)


Range of the Stop Flow Rate is wrong 

The Range of the Stop Flow Rate for several channels isn't correct.  The wrong value we're appearing in the Liquid Class Editor: Liquid Details and Help. Unfortunately, a wrong value in VENUS one and two will be abort the method.

As a workaround, please use following values:

5 ml Channels:               10-2500 ul/s
384 CO-RE Head 50 ul:   1.5-240 ul/s