Low Volume Filter Water Dispense Surface (Microlab STAR Liquid Editor)


Low Volume Filter Water DispenseSurface: Max. pipette volume to high

The Value of the max. pipette volume is with 15ul to high. The max. pipette volume is 10ul.

As a Workaround, please use following instructions:

Rename the liquidclass and then set the the value for 10ul (Target)  to 10ul (Corrected). Delete the value for 15ul (Target and Corrected). To use the LowVolumeFilter_Water_DispenseSurface_Empty LC is also possible.