HF applications containing oxides destroy normal pH glass when pH values are low. HAMILTON “HF“ pH glass
withstands HF better than other pH glasses. This significantly extends sensor life.
pH: 0 - 14
-10 - 100 °C
Max. 6 bar at 100 °C
For < 0.01M / 00 mg/l HF at 0 °C
or 0.05M / 1000 mg/l HF at 50 °C
• POLISOLVE Electrolyte
• HF applications containing oxides destroy normal pH glass when pH values are low
• HAMILTON “HF“ pH glass withstands HF better than other pH glasses. This significantly extends sensor life.
• Stable measurement values under tough conditions
• SINGLE PORE‘s for interference free contact of POLISOLVE electrolyte with measurement medium
• Pt100 in the VP version
• Serialized with batch number
• Test Certificate contains millivolt measurements in pH buffers