PRP-X500 Guard Columns

PEEK guard columns help protect your investment.

Roll the cursor over the menu headers below to see the available selections. Filter by one or more items in any order to narrow your search down to a single part number in this category. Reset a filter by clicking the red X that appears next to a filtered item. For more detail on any menu item, click the question mark on the right-hand side of the menu bar.

              Part Number
              Description
              Unit Price
              Quantity

              Hamilton guard columns protect your analytical HPLC column. They remove particulate contaminants and highly adsorptive compounds from samples, prolonging analytical column life.

              Guard columns are modular and cartridge replacement is easy and tool free. Guard column cartridge holders are available in PEEK. The design is optimized to reduce dead volume and prevent extra column band broadening.

              Holders for the cartridges are reusable indefinitely. Technical specifications and an exploded drawing are shown below for the guard column and holder. As with all guard columns these should be installed before the analytical column and after the injection valve.

               

              Guard Columns for PEEK HPLC Columns

               



               

               Product Specifications

              Housing Material:

              Polyetheretherketone (PEEK) and Delrin

              Connecting Tubing Size and material:

              1/16" O.D., 0.25mm I.D., PEEK

              Cartridge Materials:

              Barrel: PEEK

              Frits: Ceramic

              Frit Porosity: 2 µm

              Cartridge Dimension:

              8.0 x 3.0mm I.D.

              Cartridge Bed Volume:

              56 µL

              Cartridge Packing Material:

              As labeled on the cartridge

              Particle Size:

              12-20µm for polymeric packings

              10 µm for silica supports

              System Pressure Rating:

              0-5,000 psi

              Chemical Compatibility:

              Recommended for use with virtually all solvents except, concentrated nitric acid, sulfuric acid, tetrahydrofuran or dimethylsulfoxide.

               

              Selecting a Guard Column Packing

              Choose a guard column cartridge with the same packing as your analytical HPLC Column.

              Conditioning The Guard Column

              After installing the guard column cartridge into the holder with the flow arrow pointing toward the analytical column, pump approximately 2 mL of mobile phase through the cartridge before connecting it to the analytical HPLC column.

              Sealing the Guard Cartridges

              The guard cartridge holder is designed to seal with finger-tight force. If the guard column leaks, loosen the holder and re-tighten. If the leak persists, remove the cartridge and inspect the cartridge and holder for signs of damage or wear. Replace any damaged or worn parts and reassemble.


              For Sales and Product information, contact your local Hamilton Sales Office at  .   Bookmark and Share