| Product |
Chemical |
Gas |
Autoclave |
Comments |
 |
| N - Cemented Needle Syringes |
Yes, no bleach |
Yes |
No |
Glass and steel expand at different rates |
 |
| LTN - Luer Tip Cemented Needle Syringes |
Yes, no bleach |
Yes |
No |
Glass and steel expand at different rates |
 |
| RN and RNFS - Removable Needles |
Yes, no bleach |
Yes |
Yes, only when disassembled |
Remove ferrule from needle |
 |
| LT - Luer Tip Syringes with no needle |
Yes |
Yes |
Yes, only when disassembled |
|
 |
| KH - 7000 Series Syringes |
Yes |
Yes |
Yes, only when disassembled |
|
 |
| CA - Carbon Analyzer Syringes |
Yes, no bleach |
Yes |
No |
Glass and steel expand at different rates |
 |
| CR - Constant Rate Syringes |
Yes, no bleach |
Yes |
No |
Glass and steel expand at different rates |
 |
| SL - Sample Lock Syringes |
Yes, no bleach |
Yes |
No |
Glass and steel expand at different rates |
 |
| AD and DAD - Accudil Syringes |
Yes, no bleach |
Yes |
No |
Glass and steel expand at different rates |
 |
| C and CX - Chem Syringes |
Yes |
Yes |
Yes, only when disassembled |
Glass and steel expand at different rates |
 |
| D and DX - Diluter Syringes |
Yes |
Yes |
Yes, only when disassembled |
Glass and steel expand at different rates |
 |
| RN - Removable Needles 22-26 Gauge |
Yes |
Yes |
Yes, only when disassembled |
Remove ferrule from needle |
 |
| RN - Removable Needles 27-33 Gauge |
Yes |
Yes |
No |
Non-removable ferrule expands at different rate than the needle |
 |
| N - Metal Hub Needles |
Yes |
Yes |
Yes |
|
 |
| KF - Kel F Hub Needles |
Yes |
Yes |
No |
Hub expands at different rate than the needle |
 |
| PTFE Tubing |
Yes |
Yes |
Yes |
|
 |
| Kel-F Fittings |
Yes |
Yes |
Yes |
|
 |