Syringe Size (µL) Percent Stroke Accuracy (±) Precision
10 5% ≤ Stroke < 30% 3.0% 2.0%
30% ≤ Stroke 2.0% 0.2%
25 5% ≤ Stroke < 30% 3.0% 2.0%
Stroke = 30% 1.5% 0.2%
Stroke = 100% 1.0% 0.2%
50 5% ≤ Stroke < 30% 2.0% 1.0%
Stroke = 30% 1.0% 0.2%
Stroke = 100% 1.0% 0.2%
100 5% ≤ Stroke < 30% 3.0% 1.5%
Stroke = 30% 1.2% 0.5%
Stroke = 100% 1.0% 0.2%
250 5% ≤ Stroke < 30% 3.0% 1.5%
Stroke = 30% 1.2% 0.5%
Stroke = 100% 1.0% 0.2%
500 1% ≤ Stroke < 5% 3.0% 1.5%
5% ≤ Stroke < 30% 1.2% 0.5%
Stroke = 30% 1.0% 0.2%
Stroke = 100% 1.0% 0.2%
1,000 & larger 1% ≤ Stroke < 5% 3.0% 1.5%
5% ≤ Stroke < 30% 1.2% 0.5%
Stroke = 30% 1.0% 0.2%
Stroke = 100% 1.0% 0.2%


Calibration Values at 100%, 50%, and 10%

Syringe Size (µL) Percent Stroke Dispense Volume (µL) Accuracy (±) Precision
10 10% 1 3.0% 2.0%
50% 5 1.0% 0.2%
100% 10 1.0% 0.2%
25 10% 2.5 3.0% 2.0%
50% 12.5 1.0% 0.2%
100% 25 1.0% 0.2%
50 10% 5 2.0% 1.0%
50% 25 1.0% 0.2%
100% 50 1.0% 0.2%
100 10% 10 3.0% 1.5%
50% 50 1.0% 0.2%
100% 100 1.0% 0.2%
250 10% 25 3.0% 1.5%
50% 125 1.0% 0.2%
100% 250 1.0% 0.2%
500 10% 50 1.2% 0.5%
50% 250 1.0% 0.2%
100% 500 1.0% 0.2%
1,000 10% 100 1.2% 0.5%
50% 500 1.0% 0.2%
100% 1,000 1.0% 0.2%
2500 10% 250 1.2% 0.5%
50% 1,250 1.0% 0.2%
100% 2,500 1.0% 0.2%
5000 10% 500 1.2% 0.5%
50% 2,500 1.0% 0.2%
100% 5,000 1.0% 0.2%
10,000 10% 1,000 1.2% 0.5%
50% 5,000 1.0% 0.2%
100% 10,000 1.0% 0.2%
25,000 10% 2,500 1.2% 0.5%
50% 12,500 1.0% 0.2%
100% 25,000 1.0% 0.2%
50,000 10% 5,000 1.2% 0.5%
50% 25,000 1.0% 0.2%
100% 50,000 1.0% 0.2%
Hamilton Company USA
4970 Energy Way
Reno, Nevada 89502

Tel: (800) 648-5950
Fax: (775) 856-7259

E-mail: sales@hamiltoncompany.com
http://www.hamiltoncompany.com

Sales/Support
1 (888) 525-2123




Copyright © 1998-2011 Hamilton Company. All Rights Reserved.        Website Feedback   |    Terms of Use   |    Privacy Policy   |    Partner Portal
For Sales and Product information, contact your local Hamilton Sales Office at (800) 648-5950.