Error Messages Code Guide
The MICROLAB®'s display windows provide you with system status and with error message information. If an error occurs, an error message appears, the system beeps three times, and all operation and function keys are disabled. To recover from the error, reinitialize the system by pressing the Run/Stop key. You cannot use the probe button on a hand probe to reinitialize the system.
Battery Errors
Communication Errors
Custom Method Errors
Delay Time Errors
Dilution Errors
Dispense Errors
Drive Unit Errors
Method Errors
Priming Errors
Syringe Errors
Titration Errors
Battery Errors
Communication Errors
 |
Comunication
The communication link to the drive unit could not be established. Check that the connection to the drive unit is good and that the protocol and baud rate settings are correct. Also try cycling power on the drive unit. Then try the operation again. |
 |
Communication invalid echo
The echo received from the drive unit was not correct. Consult the user's manual to ensure the cable, loopback switch, protocol switch, and baud rate switches are set correctly. |
 |
Communication invalid length
The drive unit responded to the last command with either too little or too much data. Consult the user's manual to ensure the cable, loopback switch, protocol switch, and baud rate switches are set correctly. |
 |
Communication receive time-out
A serial I/O character was not received within the expected amount of time. Consult the user's manual to ensure the cable, loopback switch, protocol switch, and baud rate switches are set up properly. |
 |
Communication transmit
A physical serial I/O error occurred while attempting to send a command to the drive unit. This problem should be reported to your distributor. |
 |
Communication invalid character
An invalid response was received from the drive unit for the last command sent. Consult the user's manual to ensure the cable, loopback switch, protocol switch, and baud rate switches are set correctly. |
 |
|
Custom Method Errors
 |
Copy method to custom method
A titrate method cannot be copied to a custom method. All other method types may be converted to a custom method. |
 |
Custom method is empty
At least one operation must be specified in a custom method before the method can be accepted. |
 |
Custom method over-aspirate
This method will not operate as expected because more volume has been aspirated or filled than will fit in the syringe. |
 |
Delay Time Errors
Dilution Errors
Dispense Errors
Drive Unit Errors
Method Errors
 |
Duplicate method name
The method name selected cannot be used because a method with the same name already exists. Use the DELETE AN EXISTING METHOD selection to remove that method. |
 |
Method memory cleared
The battery used to preserve methods stored on this unit is good, but any stored methods have been lost. This can be caused by battery replacement or by the
first power-up of the unit. |
 |
Method memory empty
The operation cannot be performed because the method memory is empty. Use the CREATE A NEW METHOD selection to create a new method. |
 |
Method memory
A new method cannot be created because the method memory is full. Use the DELETE AN EXISTING METHOD selection to release method memory. |
 |
Priming Errors
 |
Prime cycles range
The number of prime cycles entered is not allowed on this unit. The valid number of prime cycles appear on the bottom of the screen while the number of prime cycles is being entered. |
 |
|
Syringe Errors
Titration Errors
|
 |